HT7005
USER MANUAL
1. PRECAUTIONS AND SAFETY MEASUREMENTS
This instrument complies with safety Standard IEC/EN61010-1
related to electronic measuring instruments. For your own safety and
to avoid damaging the instrument follow the procedures described in
this instruction manual and read carefully all notes preceded by this
symbol
CAUTION
Avoid measuring in humid or wet places
Avoid measuring in rooms where explosive
gas, combustible gas, steam or excessive dust
is present
Keep you insulated from the object under test
Do not touch exposed metal parts such as test
lead ends, sockets, fixing objects, circuits etc.
Avoid doing that if you notice anomalous
conditions such as breakages, deformations,
fractures, leakages of battery fluid, blind display
etc.
Be careful when measuring on circuits with
voltages exceeding 20V to avoid risks of
electrical shock
The following symbols are used in user manual and on the meter:
CAUTION - refer to the instruction manual - an
improper use may damage the instrument or its
components
Double insulated meter
This symbol indicates that the clamp can operate on
live conductors
AC and DC current
Ground reference
CAUTION: this symbol indicates that equipment, its
accessories and battery shall be subject to a
separate collection and correct disposal
1.1 Preliminary instructions
CAUTION
This instrument has been designed for use in
environments with pollution degree 2
It can be used for AC and DC CURRENT
measurements on installations with
measurement category CAT III 300V
Do not test or connect to any circuit exceeding
the specified overload protection
Do not effect measurements under
environmental conditions exceeding the limits
indicated in § 11.2
Make sure that battery is properly installed
When the instrument is connected to
measuring circuits never touch any unused
terminal
If during measurement the displayed values
remain constant check whether the HOLD
function is active
1.2 Measuring overvoltage category definitions
According to the IEC/EN61010-1 guidelines the circuits are divided
into the following measurement categories:
CAT IV is for measurements performed at the source of the low-
voltage installation
Examples are electricity meters and measurements on primary
overcurrent protection devices and ripple control units
CAT III is for measurements performed in the building installation
Examples are measurements on distribution boards, circuit
breakers, wiring, including cables, bus-bars, junction boxes,
switches, socket-outlets in the fixed installation, and equipment
for industrial use and some other equipment, for example,
stationary motors with permanent connection to fixed installation
CAT II is for measurements performed on circuits directly
connected to the low voltage installation
Examples are measurements on household appliances, portable
tools and similar equipment
CAT I is for measurements performed on circuits not directly
connected to MAINS
Examples are measurements on circuits not derived from
MAINS, and specially protected (internal) MAINS-derived circuits.
In the latter case, transient stresses are variable; for that reason,
the standard requires that the transient withstand capability of the
equipment is made known to the user
2. GENERAL DESCRIPTION
The instrument has the following features:
DC/AC TRMS Current measurement up to 400A in Autorange
Inrush current measurement
Data HOLD and ZERO features
Backlight
Detection of AC Voltage without contact
Auto Power OFF
The selected quantity is displayed with indication of measuring
unit and active functions. The instrument is provided with an Auto
Power Off function consisting in an automatic switching off 15
minutes after last pressure on keys.
3. PREPARATION FOR USE
3.1 Initial checks
This instrument was checked both mechanically and electrically
prior to shipment. All possible cares and precautions were taken
to let you receive the instrument under perfect conditions.
Notwithstanding we suggest you to check it rapidly (any damage
may have occurred during transport – if so please contact the
local distributor from whom you bought the item).
Make sure that all standard accessories mentioned in § 11.2 are
included. Should you have to return back the instrument for any
reason please follow the instructions mentioned in § 12
3.2 Supply voltage
The instrument is powered by 2x1.5V battery type AAA IEC LR03
included on meter. When battery is low the symbol "
"is
displayed. To replace/insert the batteries follow the instructions
indicated in § 8
3.4 Storage
After a period of storage under extreme environmental conditions
exceeding the limits mentioned in § 11.2 let the instrument
resume normal operating conditions before using it
4. INSTRUMENT AND DISPLAY DESCRIPTION
5. SAFETY LINE
Put the cable as close as possible to the middle of the jaws
(considering the alignment marks) in order to meet the meter
accuracy specifications and take the hand under the safety line (see
below figure)
Copyright HT ITALIA 2017 Release EN 1.00 - 27/02/2017
6. FUNCTION KEYS DESCRIPTION
6.1
key
Press
key to switch on the instrument. Press and hold the key
to switch off the instrument
6.2 DH/
key
By pressing DH/
key the measured value is hold on the display
where the symbol "H" appear. Press again the key to disable this
function and resume normal operation. Press and hold the
DH/
key to activate/deactivate the display backlight
6.3 MODE key
The MODE key allows the selection of the type of current
measurement. The “AC” and “DC” symbols are shown at display
6.4 ZERO key
The ZERO key, which is active only in DC current measurement,
allows to activate/deactivate the zeroing of value at display in
order to remove the residual magnetization (see § 7.1) ). The “
”
symbol is shown at display
6.5 INRUSH key
The INRUSH key, which is active only in AC current measurement,
allows to activate/deactivate the Inrush current measurement
(instantaneous current peaks) of a installation. The “INRUSH”
symbol is shown at display and the 400A range is automatically
selected. Press and hold the INRUSH key to select the 40A range.
Press INRUSH key again to exit by the function
6.6 Detection of AC Voltage without contact
1. Switch on the instrument with
key
2. Move the NCV sensor (see figure) close to AC source
3. The switch on of red LED indicate the voltage presence
6.7 Auto Power OFF feature
In order to preserve internal batteries, the instrument switches
automatically off approximately 15 minutes after it was last used.
The “
” symbol appears on the display when this function is
active
7. HOW TO PERFORM THE MEASUREMENTS
7.1 DC and AC Current measurement
CAUTION
Put the conductor as close as possible to the
middle of the jaws in order to meet the meter
accuracy specifications
When apply/remove the instrument on/from live
conductors, if is not possible to de-energize the
circuit on test, use appropriate security
measures (e.g. protective gloves) to avoid
dangerous electric shock for the operator and a
possible damage of the instrument
Disconnect the main voltage or wear protective
gloves before performing measurement
Put the hand below the safety line during the
measurement as indicated in § 5
1. Switch on the instrument by pressing key
2. Press MODE key to select AC or DC measurement
3. For DC current measurement press the ZERO key in order to
remove the residual magnetization
4. Open the clamp jaw and insert the cable in the middle of it (see
figure – left part). Take care to the polarity for DC current
measurement
5. Do not perform measurements with more cables included in the
clamp jaw (see figure – right part)
6. The AC or DC current value will be shown on the display
7. If the message "OL" is displayed the maximum current value was
exceeded
8. For Data HOLD and INRUSH features see § 6.2 and § 6.5
8. BATTERY REPLACEMENT
When the "
" low battery indication is displayed the battery must
be replaced.
CAUTION
Only expert technicians should perform this operation.
Before carrying out this operation, make sure you
have removed the cable being tested from inside the
clamp jaw
1. Switch off the instrument by pressing
key
2. Remove the fixing screw and the battery compartment cover
3. Remove the batteries and replace them with new one of the same
type (see § 11.2) by respecting the correct polarity
4. Replace battery compartment cover
5. Do not scatter old batteries into the environment. Use the relevant
containers for disposal
9. CLEANING THE INSTRUMENT
Use a soft and dry cloth to clean the instrument. Never use wet cloths,
solvents, water, etc.
10. END OF LIFE
CAUTION: this symbol indicates that equipment, its
accessories and battery shall be subject to a
separate collection and correct disposal
11. TECHNICAL SPECIFICATIONS
11.1 Technical characteristics
Accuracy declared at temperature: 23°C 5°C, <80%RH
DC/AC TRMS Current
For not sinusoidal waveforms the frequency range is: 50Hz ÷60Hz
Overload protection: 400A AC
INRUSH current 40A range: 3A ÷ 40A, 50/60Hz, Accuracy: 10%rdg
INRUSH current 400A range:10A ÷ 400A, 50/60Hz, Accuracy: 10%rdg
INRUSH current response time: 100ms
11.2 General specifications
Reference guidelines
Safety: IEC/EN61010-1, EMC: IEC/EN61326-1
Measurement category: CAT III 300V
Pollution degree: 2, Insulation: double insulation
Max height of use: 2000m (6562ft)
Mechanical protection: IP30
Compliance with 2014/30/EU (EMC) and 2014/35/EU (LVD)
Display
Characteristics: LCD, 3 ¾dgt, 4000 points + decimal point
Conversion type: TRMS
Sampling rate: 3 times/s
Over range indication: “OL” symbol at display
Power supply
Battery type: 2x1.5V battery type AAA IEC LR03
Battery life: ca 85h (with backlight), ca 240h (without backlight)
Low battery indication: symbol “
” at display
Auto Power OFF: after 15 minutes of idleness
Environmental conditions for use
Working temp.: 540°C (41°F 104°F), <80%RH
Storage temperature: -20 60°C (-4°F 140°F), <80%RH
Mechanical characteristics
Dimensions (L x W x H): 155x60x25mm (6x2x1in)
Weight (with batteries): 140g (5 ounces)
Max diameter cable: 20mm (1in)
Standard accessories
Batteries
Carrying case
User manual
Range Resolution Frequency Accuracy
40.00A AC 0.01A
50Hz60Hz
(2.5%rdg+10dgt)
400.0A AC 0.1A
(2.8%rdg+8dgt)
40.00A AC 0.01A
60Hz400Hz
(2.8%rdg+10dgt)
400.0A AC 0.1A
(2.8%rdg+8dgt)
40.00A DC 0.01A
DC
(2.5%rdg+10dgt)
400.0A DC 0.1A
(2.8%rdg+8dgt)
12. SERVICE
12.1 Warranty conditions
This instrument is guaranteed for one year against material or
production defects, in accordance with our general sales conditions.
During the warranty period the manufacturer reserves the right to
decide either to repair or replace the product.
The warranty shall not apply in the following cases:
Repair and/or replacement of accessories and battery (not
covered by warranty).
Repairs that may become necessary as a consequence of an
incorrect use of the instrument or due to its use together with
non-compatible appliances.
Repairs that may become necessary as a consequence of
improper packaging.
Repairs which may become necessary as a consequence of
interventions performed by unauthorized personnel.
Modifications to the instrument performed without the
manufacturer’s explicit authorization
Use not provided for in the instrument’s specifications or in the
instruction manual
12.2 Service
Should the instrument not work properly, make sure that battery is
correctly installed and working and replace if necessary before
contacting your distributor. Should you need for any reason to return
back the instrument for repair or replacement take prior agreements
with the local distributor from whom you bought it. Do not forget to
enclose a report describing the reasons for returning (detected fault).
Use only original packaging. Any damage occurred in transit due to
not original packaging will be charged anyhow to the customer. The
manufacturer will not be responsible for any damage to persons or
things
HT ITALIA SRL
Via della Boaria, 40
48018 – Faenza (RA) – ITALY
Tel: +39-0546-621002
Fax: +39-0546-621144
Web: www.ht-instruments.com
ENG

Chatten Sie mit Ihrem HT Instruments HT7005-Handbuch mithilfe von KI

Erhalten Sie sofortige KI-Antworten aus Ihrem HT Instruments HT7005-Handbuch

Fragen Sie alles über HT Instruments HT7005. Antworten stammen aus diesem Handbuch, mit den Seiten, auf denen sie basieren.

HT7005 Wartungsplan

Bei Anzeige
Batterie ersetzen

Batterien ersetzen, wenn das Symbol für niedrigen Batteriestand auf dem Display erscheint. Verwenden Sie zwei 1,5-V-AAA-Batterien unter Beachtung der Polarität.

Bei Bedarf
Gerät reinigen

Verwenden Sie ein weiches, trockenes Tuch zum Reinigen des Geräts. Verwenden Sie keine feuchten Tücher, Lösungsmittel oder Wasser.

M: QW-35-FL

Stellen Sie eine Frage

0/1000

Empfehlen Sie HT7005?

Ja Nein

Häufig gestellte Fragen

A

Das HT7005 misst AC/DC TRMS-Ströme bis 400 A mit automatischer Messbereichswahl. Beispielgenauigkeiten: 40,00 A AC (50–60 Hz) ±(2,5 % des Messwerts + 10 Digits); 400,0 A AC ±(2,8 % des Messwerts + 8 Digits). Die INRUSH-Genauigkeit beträgt ±10 % für die 40 A- und 400 A-INRUSH-Bereiche.

Ist diese Antwort hilfreich?

A

Ja. Die INRUSH-Funktion erfasst transiente Anlaufspitzen mit dem automatischen 400 A-Bereich; zum Auswählen des 40 A-Bereichs drücken und halten. Die INRUSH-Reaktionszeit beträgt etwa 100 ms.

Ist diese Antwort hilfreich?

A

Es verwendet 2 × 1,5 V AAA (IEC LR03) Batterien. Die typische Batterielaufzeit beträgt etwa 85 Stunden mit Hintergrundbeleuchtung und ca. 240 Stunden ohne Hintergrundbeleuchtung. Ein Batterie-Warnsymbol erscheint auf der Anzeige, wenn ein Austausch erforderlich ist.

Ist diese Antwort hilfreich?

A

Ja, bei korrekter Verwendung. Das Messgerät ist doppelt isoliert, CAT III 300 V bewertet und konform mit IEC/EN61010-1. Positionieren Sie stets den Leiter in der Zangenmitte, halten Sie die Hand unter der Sicherheitsreferenz und befolgen Sie PSA-Richtlinien; vermeiden Sie Messungen in nassen, explosionsgefährdeten oder stark staubigen Umgebungen.

Ist diese Antwort hilfreich?

A

Der Betriebstemperaturbereich liegt bei 5–40 °C mit <80% relativer Luftfeuchte; Lagerung −20 bis 60 °C. Maximale Höhe 2000 m. Mechanischer Schutz ist IP30 und die Zange akzeptiert bis zu 20 mm Kabeldurchmesser.

Ist diese Antwort hilfreich?

A

Verwenden Sie ein weiches, trockenes Tuch zur Reinigung. Verwenden Sie kein Wasser, keine Lösungsmittel und keine feuchten Tücher. Entfernen Sie geprüfte Leiter, bevor Sie Batterien wechseln. Lagern Sie das Messgerät innerhalb der angegebenen Umgebungsgrenzen und lassen Sie es nach extremen Lagerbedingungen vor dem Gebrauch wieder normalisieren.

Ist diese Antwort hilfreich?

A

Standardzubehör umfasst Batterien und ein Trageetui. Am Lebensende müssen das Gerät, Zubehör und die Batterien getrennt entsorgt und gemäß den örtlichen Vorschriften für Elektro- und Batterieabfall behandelt werden.

Ist diese Antwort hilfreich?

M: OA-GPT-5-NAN

Vorteile und Nachteile

Vorteile

  • Präzise TRMS-AC/DC-Strommessung bis 400 A mit automatischer Messbereichswahl.
  • INRUSH-Modus erfasst Anlauftransienten mit wählbaren 40 A/400 A-Bereichen.
  • NCV-berührungsfreie Spannungsdetektion und hintergrundbeleuchtetes LCD für schnelle Feldprüfungen.
  • Kompaktes, leichtes Design (140 g) mit langer Batterielaufzeit und automatischer Abschaltung.
  • Sicherheitsbewertet nach IEC/EN61010-1, doppelte Isolierung, CAT III 300 V, IP30.
VS

Nachteile

  • Begrenzte Zangenöffnung (20 mm) schränkt die Verwendung an gebündelten oder großen Leitern ein.
  • IP30 bietet eingeschränkten Staub-/Wasserschutz – ungeeignet für raue, nasse Umgebungen.
  • INRUSH-Genauigkeit ist ±10 %, was für präzise Transientenanalysen grob sein kann.
  • Batteriezugang empfohlen nur für erfahrene Techniker.
  • Keine erweiterte Protokollierung oder PC-Schnittstelle für umfangreiche Datenanalyse.
M: OA-GPT-5-NAN

Produkt Spezifikation

Messbereich

max. 400 A, max. 300 V

Lieferumfang

Tasche, Batterie

M: GO-GEM-FL-25

Produkt Beschreibung

HT Instruments HT7005 Stromzangen-Multimeter Handbuch Übersicht

Das HT Instruments HT7005 Stromzangen-Multimeter ist ein kompaktes, automatisch messbereichswählendes Zangenmessgerät, das präzise AC/DC TRMS-Strommessungen bis 400 A ermöglicht und zusätzliche Funktionen für Feld- und Laborarbeiten bietet. Es kombiniert ein sicherheitsorientiertes Design — konform mit IEC/EN61010-1 und bewertet als CAT III 300 V — mit praktischen Funktionen wie Erfassen von Einschaltströmen (Inrush), Datenhaltefunktion, Nullstellung zur Entfernung von DC-Vorspannung und berührungsfreier AC-Spannungsdetektion. Sein leichtes Gehäuse, das hintergrundbeleuchtete LCD und die lange Batterielaufzeit machen es geeignet für Elektriker, HLK-Techniker und Instandhaltungspersonal.

Hauptmerkmale und Nutzung

Das HT7005 liefert TRMS-Stromwerte für sinusförmige und nicht-sinusförmige Wellenformen in den angegebenen Frequenzbereichen und verfügt über eine automatische Messbereichswahl für die 40 A- und 400 A-Skalen. Die INRUSH-Funktion erfasst transiente Anlaufströme (wählbare 40 A/400 A-Bereiche) mit einer Reaktionszeit von ca. 100 ms. Für Gleichstrommessungen kompensiert die ZERO-Funktion Restmagnetisierung zur Verbesserung der Genauigkeit bei niedrigen Strömen. Der NCV-Sensor (berührungsfreie Spannungsprüfung) und die rote LED ermöglichen schnelle Prüfungen auf AC-Spannung von etwa 200 V bis 1000 V.

Sicherheit und korrekte Handhabung werden betont: Positionieren Sie den Leiter in der Mitte der Zange für beste Genauigkeit, halten Sie die Hände unterhalb der Sicherheitsreferenz und vermeiden Sie Messungen in nassen oder explosionsgefährdeten Umgebungen. Das Messgerät verfügt über Überlastschutz bis 400 A AC, Doppelte Isolierung und mechanischen Schutz nach IP30. Betrieben wird es mit zwei AAA-Batterien; das HT7005 verfügt über automatische Abschaltung zur Energieeinsparung und zeigt ein Batterie-Warnsymbol an, wenn ein Austausch erforderlich ist.

Praktische Hinweise

Das Gerät wird mit Batterien und einem Trageetui geliefert, wiegt etwa 140 g mit Batterien und akzeptiert Leiter mit einem Durchmesser bis 20 mm. Typische Betriebsbedingungen liegen bei 5–40 °C, und es ist für den Einsatz bis 2000 m Höhe ausgelegt. Genauigkeitsangaben variieren je nach Bereich und Frequenz (z. B. 40,00 A AC bei 50–60 Hz: ±(2,5 % des Messwerts + 10 Digits); 400,0 A AC: ±(2,8 % des Messwerts + 8 Digits)). Befolgen Sie empfohlene Sicherheitsverfahren bei Messungen an unter Spannung stehenden Leitern und vermeiden Sie das Einführen mehrerer Leiter in die Zange.

Hinweise

  • Stellen Sie sicher, dass die Batterien mit der richtigen Polarität eingesetzt sind und ersetzen Sie beide AAA-Zellen bei Bedarf; nur qualifiziertes Personal sollte das Batteriefach öffnen.
  • Reinigen Sie mit einem weichen, trockenen Tuch; vermeiden Sie Lösungsmittel und Feuchtigkeit.
  • Am Ende der Lebensdauer sind Gerät und Batterien entsprechend den örtlichen Vorschriften für Elektro- und Batterieabfall zu entsorgen.